The YG6100 X-ray fluorescence sulfur analyzer is a new type of computerized desktop instrument that uses pure physical analysis methods. It can detect the percentage content of SO3 in cement, gypsum, etc. within 30 seconds,
Compared with cement chemical analysis method and gypsum chemical analysis method, which are fast and accurate, by quickly determining the percentage content of SO3 in cement, cement enterprises can adjust gypsum in cement in real time
The addition amount and prediction of the setting time of concrete produced with this type of cement can also be applied to other occasions where SO3 percentage content needs to be analyzed, such as detecting desulfurization ash in power plants
The percentage content of SO3.
Large screen LCD display, full Chinese menu prompts for operation, intelligent error alarm prompts, extremely convenient to use.
Fast analysis speed, able to detect the percentage content of SO3 in cement and gypsum in 30 seconds.
No need for any chemical reagents, no discharge of three wastes, no radioactive sources, low power consumption, in line with environmental protection, energy conservation and safety requirements.
Main technical indicators:
- Analysis range: The SO3 analysis range can be adjusted and selected by calibrating the working curve.
- Analysis range width: SO3% max – SO3% min ≤ 5%;
For example, cement SO3: 0.1~5%. - Inherent error: SO3: ± 0.11%.
- System analysis time: 30 × n seconds (n is a natural number), recommended value is 60 seconds.
- Analysis accuracy: SO3 ≤ 0.05%.
- Temperature stability: Within the range of 5~+40 ℃, absolute drift: ∣ △ SO3% ∣ ≤ 0.11%.
- Usage conditions: Environmental temperature: 5~+40 ℃, relative humidity: ≤ 85% (30 ℃), power supply: 220V ± 20V, 50Hz.
- Power consumption of the whole machine:<30Kw
- Weight: 468 × 368m × 136m, 13.8Kg