The HT-H2000-2 is a fully automatic instrument for measuring coke reactivity and post-reaction strength, featuring a three-zone dual-control reactivity testing system.
Features:
- Utilizes imported high-temperature heating elements with an integrated electromechanical design. The reactor loading/unloading mechanism employs bottom-mounted mechanical transmission for easy installation and operation.
- The control system is based on Siemens PLC technology, with gas flow regulated by mass flow controllers (MFCs) for precise and real-time flow measurement and display. Features self-learning temperature control algorithms, programmable settings, and multiple control modes including open-loop stepwise and closed-loop fully automatic.
- Temperature transmitters use two-wire solid-state modules with automatic temperature compensation and linearization.
- Capable of simulating the entire testing process to verify system performance.
- Automatically activates the ventilation system during testing to expel indoor CO gas, ensuring operator safety and air circulation.
- Connects to a touchscreen computer for real-time display of temperature and gas flow curves. Includes a CO2 flow deficiency alarm.
- Equipped with over-temperature alarm and automatic shutdown to prevent damage from thyristor failure, thermocouple damage, or signal interference.
- Includes over-pressure protection.
- Thermocouples are protected by GH3030 high-temperature alloy steel sleeves.
Technical Parameters:
A. Standard Control Process:
- Heating from room temperature to 1100°C at a rate of 15°C/min (adjustable).
- Introduces 0.8L/min N2 for protection at 400°C.
- Preheats CO2 at 1050°C.
- Stabilizes at 1100±5°C for 10min, then switches to 5L CO2 for a 2h reaction.
- Furnace stops heating; reactor descends and exits the furnace. CO2 is cut off, and 2L N2 is introduced for protection.
- An alarm sounds when the coke cools to 100°C, marking the end of the test.
- The entire process is automatically controlled by a Siemens PLC.
B. Temperature Control:
- Three-point thermocouples: Type S or N, Class 0.5; control accuracy: 1100°C±3°C.
- Protection tube: Alumina.
- Heating program: Room temperature to 1100°C at 15°C/min; holds at 1100°C for 2h.
C. Flow Control:
- Automated mass flow controller (MFC); includes 2 drying tubes.
- N2: 0-10L adjustable; CO2: 0-10L adjustable.
- Accuracy: ±1.5% F.S.
D. Reactor:
- Material: GH3044; maximum operating temperature: ≤1400°C.
- Reactor dimensions: Inner diameter Φ80mm.
E. Type I Drum:
- Dimensions: Φ140mm×700mm; wall thickness: 5mm.
- Rotation: 20±0.5 rpm, automatically controlled for 600 rotations with automatic counting.
F. Test Sieves:
- Square-hole sieves (optional, ISO standard): 19mm, 22.4mm, 10mm.
- Round-hole sieves: 23mm, 25mm, 10mm.
G. Furnace:
- Integrated furnace chamber; rated power: 7.5KW.
- Power supply: Three-phase five-wire system with balanced load.
- Three independently controlled heating zones with a 300mm uniform temperature zone. Displays upper, middle, and lower temperature control curves on the touchscreen.
- Maximum operating temperature: ≤1250°C.
- Dimensions: 1000mm×850mm×1800mm.
- Total weight: <500kg.