The HT-2000 Coke Reactivity and Post-Reaction Strength Tester is an instrument used to measure the reactivity index of coke. This equipment employs PLC control and an industrial touchscreen for display. The high-temperature furnace’s temperature is regulated using closed-loop fuzzy logic PID control, while the PLC system comprehensively monitors and processes the central temperature of the furnace chamber. The touchscreen allows for operation and modification of PLC parameters, with dynamic display of variable curves.
I. Working Environment and Installation Conditions:
- The electrical control system adopts TN-S wiring.
- The electrical equipment casing must be reliably grounded.
- The electrical equipment must be installed in an environment free from vibration, dust, and particulate matter.
- Operating voltage: AC220V±10%, 50Hz±5%, three-phase five-wire system.
- Altitude ≤ 2000m.
- Operating environment: 5℃ ≤ ambient temperature ≤ 40℃, humidity ≤ 85%, no strong magnetic fields, and no severe vibrations.
II. Technical Performance and Parameters:
- Dynamic display of the operational status of all system components.
- Industrial touchscreen display.
- Dynamic display of the sidewall temperature values of the high-temperature heating furnace.
- Dynamic display of the central temperature curve of the high-temperature heating furnace.
- Dynamic display of the gas flow curve in the gas path.
- Dynamic display of the temperature setpoint of the high-temperature heating furnace.
- Dynamic display of the power output value of the high-temperature heating furnace.
- Dynamic display of the voltage output value of the high-temperature heating furnace.
- Dynamic display of the current control value of the high-temperature heating furnace.
- Dynamic display of the heating rate of the high-temperature heating furnace.
- Protection functions for excessive or insufficient heating rates.
- Over-temperature protection function.
- Gas flow interruption fault protection function.
- Thermocouple fault protection function.
- High-temperature furnace: 7KW
- Temperature control accuracy: 1100℃±5℃
- Temperature control range: 0~1150℃, resolution 1℃.
- Temperature sensor: Type N thermocouple
- Maximum operating temperature: ≤1150℃
- Heating element: Silicon carbide rod
- Gas flow accuracy: ±1.5% F.S.
- Gas flow linearity: ±(0.5~1.5)% F.S.
- Gas flow repeatability: ±0.2% F.S.
- Gas flow response time: 1~4 seconds.
- Operating voltage: AC220V±10%, 50Hz±5%, three-phase five-wire system.
- Operating environment: 5℃ ≤ ambient temperature ≤ 40℃, humidity ≤ 85%, no strong magnetic fields, and no severe vibrations.