Anti-emulsibility refers to the speed at which oil separates from water when they come into contact. If the anti-emulsibility of the lubricating oil is poor, it will form an emulsion with the mixed water, making it difficult for the water to be drained from the bottom of the circulating oil tank, which may result in poor lubrication.
This instrument complies with GB/T7305 “Determination of Water Separability of Petroleum and Synthetic Fluids” and GB/T7605 “Determination of Demulsibility of In-Service Turbine Oils”. It is mainly used to determine the ability of petroleum and synthetic fluids to separate from water. With a reasonable structure, stable performance and simple operation, it is an ideal analytical testing equipment.
Instrument Features:
1. LCD Chinese display human-machine interface, for parameters such as preset temperature and test time; with menu-prompted input;
2. It adopts microcomputer control and PID self-tuning temperature control technology, featuring fast temperature rise and high temperature control accuracy;
3. The instrument features automatic stirring, automatic timing, and automatic lifting of the large arm of the test tube stirring motor, which reduces the labor intensity of operators;
4. It automatically alarms when the test time is up and can separate 3 samples at the same time, which improves work efficiency;
5. The mechanical transmission is noiseless, stable and reliable. Internal components such as the main heater and auxiliary heater are made of stainless steel, which is corrosion-resistant and durable;
6. The constant temperature bath is a circular glass cylinder with a uniform temperature distribution inside, and the temperature control effect is good;
Technical indicators:
Display mode: | Blue screen (240*128 dot matrix) |
Temperature control range: | room temperature—100℃ |
Temperature control accuracy: | ±0.1℃ |
Timing range: | Can be set arbitrarily from 0 to 99 minutes |
Stirring speed: | 1500r±15r/min (Sample stirring) |
Temperature sensor: | Platinum resistor (imported Class A PT100) |
Number of experimental wells: | 3 |
Power supply voltage: | AC220V±20% 50HZ±10% |
Maximum power: | 1000W |
Applicable ambient temperature: | 5℃~40℃ |
Applicable ambient humidity: | ≤85% RH |
Overall dimensions: | 540×380×600(mm) |