The determination principle of this instrument is the world-recognized classic method for water measurement, the “Karl Fischer Coulometric method”. This method features high testing accuracy and extremely low testing cost. It is the most specific and accurate chemical method and also a globally recognized industry standard analysis method. The liquid can be quickly determined within one minute. Solid. The moisture content in the gas is directly displayed as the measurement result, with a measurement range of 0.0001% (1ppm) to 100%. It is widely used in the petroleum, chemical, power, pharmaceutical, pesticide, railway industries and scientific research units in colleges and universities.
The sample determination process is automatically controlled by the instrument. Stirring and determination are completed automatically within about 60 seconds, and the determination results are directly displayed. The fully enclosed titration system avoids the harm of reagents to the operators. The 500ml Karl Fischer reagent that comes with it can continuously measure about 1000 samples and has other significant advantages. It completely avoids the excessive errors that occur in the process of determining water content by other methods, thereby improving the quality indicators of the product.
Meet the standards:
GB6283-86 “Determination of Moisture Content in Chemical Products – Karl Fischer Method (General Method)”
GB/T606-2003 “Chemical Reagents – General Method for Moisture Determination – Karl Fischer Method”
GB/T11133-1989 “Method for Determination of Water Content in Liquid Petroleum Products (Karl Fischer Method)
SH / T0246 GB / T11133-1989 GB / T7380-1995 GB10670-1989 GB / T3776.1-1983
GB / T6023-1999、GB / T1600.GB7600.B/ T 18619.1.SH / T 0255
Application field:
Liquid | Chemical industry: alcohols, ethers, esters, acids, alkanes, benzene, amines, organic solvents, phenols Organic products suitable for determination by the Karl Fischer method, etc. Petroleum and power industry: insulating oil, transformer oil, turbine oil and other oil products Pharmaceutical industry: raw materials for medicine, etc Pesticide industry: emulsifiers, etc Other industries: lithium battery electrolyte, etc |
Solid | All kinds of inorganic salts, citric acid, explosives, paraffin and other solids with good solubility |
Gas | Natural gas, liquefied gas, Freon, butadiene, chloromethane and other gases |
Main features:
- It adopts A powerful new-generation microprocessor with a dual-CPU design, and the measurement electrode data can be directly converted through A/D sampling.
- 7-inch touch screen Chinese LCD display, reagent expiration prompt function, extending the service life of reagents;
- The fully sealed titration cell bottle prevents reagents from coming into contact with people and also avoids the influence of environmental humidity.
- The blank current microprocessor automatically controls compensation, and the reagent can quickly reach a balanced state.
- With a large electrolytic current of 0-430 milliampere, it features high detection sensitivity and fast analysis speed.
- Content calculation (3 formulas), statistical calculation (average value, standard deviation, change rate), and four arithmetic operations;
- The instrument automatically measures the sample, displays the analysis results automatically, and can automatically calculate the average value of multiple measurements of the sample.
- The test results are automatically stored and historical records can be accessed at any time.
- Automatically calculate and print contents such as ug/ppm/percentage content/mg/L/sample number/tester/experiment date
Main technical indicators:
Display mode | 7-inch color touch screen with Chinese LCD display |
Measurement method | Microprocessor-controlled electrolysis (Karl Fischer Coulometric method) |
Measurement range | 0.01 micrograms of water -200 milligrams of water (0.01 micrograms -9999.99 micrograms) |
Water content range | 0.0001%(1ppm)—100% |
Sensitivity | 0.01 micrograms of water |
Finish line indication | Screen display/print output/sound warning/finish light indication |
Accuracy | When the water content is between 2 micrograms of water and 100 micrograms, the error is ≤±1 microgram. When the water content is between 100 micrograms of water and 1000 micrograms, the error is ≤±2.9 micrograms. When the water content is above 1000 micrograms of water, the error is ≤±0.2%(excluding injection error) |
Electrolysis rate | 2.4 milligrams of water per minute |
Electrolytic current | 0-430 milliampere automatic control |
Drift compensation | Microprocessor automatic control |
mixing speed | Adjustable control |
Data input | Touchscreen operation input |
Sample number | User-defined |
Storage function | The result is stored infinitely. |
Printer | Low-energy consumption high-speed thermal printer, 44mm paper width |
打印内容 Print content | Ug / ppm/ mg/L /Sample number / Tester / Date |
Calendar/Clock | Analyze the time and date display and print them out |
Self-check function | Automatic diagnosis of instrument faults |
Overall dimensions | 390×270×190mm |
Usage environment | Temperature: 5℃-40℃, humidity: < 85% |
Power consumption | <40W |
Power supply | AC220V±10% 50HZ±2.5HZ |
Weight | About7kilograms |
Configuration
Attachment Name | Quantity/Unit | |
1 | Host | 1 unit |
2 | Electrolytic cell bottle | 1 |
3 | Electrolytic electrode | 1 |
4 | Measuring electrode | 1 |
5 | Drying tube | 1 root |
6 | Sealing gasket | 10 |
7 | Vacuum grease | 1 box |
8 | Power cord | 1 article |
9 | Magnetic stirring rod | 1 |
10 | 0.5 microliter sampler | 1 stick |
11 | 50 microliter sampler | 1 stick |
12 | 100 microliter sampler | 1 stick |
13 | 1 milliliter injector | 1 stick |
14 | Karl Fischer reagent | 1 bottle |
15 | Instruction Manual | 1 book |
16 | Attached | 1 book |
17 | Certificate of Conformity | 1 sheet |