The HTCS-600 micro moisture analyzer is based on the world-recognized classic water measurement method, the Karl Fischer coulometric method. It has a measurement range as low as 0.0001% and can quickly determine the liquid within 60 seconds. Solid. The moisture content in gases is widely applied in industries such as chemical engineering, petroleum, power, scientific research, and railways.
Meet the standards:
GB/T 6283-2008 Determination of moisture content in chemical products – Karl Fischer method
ASTM E1064-2008 Determination of water content in organic liquids by Karl Fischer coulometric titration
GB/T 7600-1987 Determination of moisture Content in Transformer Oil in operation (Coulometric method)
ASTM D4928-00(2010) Determination of water content in crude oil by Karl Fischer coulometric titration
ASTM D6304-2007 Karl Fischer coulometric titration is used to determine the water content in petroleum products, lubricating oils and additives
ISO 10337-1997 Determination of moisture content in crude oil by Karl Fischer coulometric titration
GB/T 11146-2009 Determination of water content in crude oil by Karl Fischer coulometric titration
GB/T 3727-2003 Determination of trace water in industrial ethylene and propylene
GB/T 5074-1985 Microcoulometric method for the determination of moisture content in coking products
GB/T 6023-2008 Determination of trace water in industrial butadiene by Karl Fischer Coulometric method
GB/T 7376-2008 Determination of trace moisture in Industrial fluorinated alkanes by Karl Fischer method
GB/T 18619.1-2002 Determination of water content in natural gas by Karl Fischer Coulometric method
GB/T 18826-2002 Industrial 1,1,1, 2-tetrafluoroethane HFC-134a
SH/T 0246-1992 Determination of water content in light Petroleum products (Electrochemical method)
SH/T 0255-1992 Determination of moisture content in additives and lubricating oils containing additives (coulometric method)
ASTM E1064-2008 Determination of water content in organic liquids by Karl Fischer coulometric titration
Main functions:
It adopts advanced microprocessor automatic control with a sensitivity of 0.1 micrograms;
The fully sealed titration cell bottle prevents reagents from coming into contact with people and also avoids the influence of environmental humidity;
The blank current microprocessor automatically controls the compensation, and the reagent can quickly reach a balanced state;
The stirring speed is electronically adjusted by the key, and there is a self-check function key for faults;
The automatic control of 400 milliampere large electrolytic current has greatly improved sensitivity and accuracy。
Main technical indicators:
Display mode | Five high-brightness digital tubes display |
Measurement method | Karl Fischer coulometric method |
Measurement range | 0.1 micrograms of water – 100 milligrams of water |
Water content range | 0.0001%(1ppm)—100% |
Sensitivity | 0.1 microgram of water |
Finish line indication | Sound warning/Finish light indication |
Accuracy | When the water content is between 1 microgram of water and 1000 micrograms, the measurement error shall not exceed ±0.2% When the water content is above 1000 micrograms of water, the measurement error is ≤0.3%(excluding injection error). |
Electrolytic current | 0-400 milliampere automatic control |
Self-check function | It is equipped with a self-check function key for faults |
Drift compensation | Microprocessor automatic control |
Overall dimensions | 300×250×100mm |
Usage environment | Temperature: 5℃-40℃ |
Power consumption | <40W |
Power supply | AC220V±10% 50HZ±2.5HZ |
Weight | About 5 kilograms |
Attachment Name | Quantity/Unit | |
1 | Host | 1 unit |
2 | Electrolytic cell bottle | 1 |
3 | Electrolytic electrode | 1 |
4 | Measuring electrode | 1 |
5 | Drying tube | 1 root |
6 | Sealing gasket | 10 |
7 | Vacuum grease | 1 box |
8 | Power cord | 1 article |
9 | Magnetic stirring rod | 1 |
10 | 0.5 microliter sampler | 1 stick |
11 | 50 microliter sampler | 1 stick |
12 | 100 microliter sampler | 1 stick |
13 | 1 milliliter injector | 1 stick |
14 | Karl Fischer reagent | 1 bottle |
15 | Instruction Manual | 1 book |
16 | Attached | 1 book |
17 | Certificate of Conformity | 1 piece |