Under specified conditions, when the oil is heated and its temperature reaches a certain level, the mixture of the oil’s vapor and the surrounding air will flash when it comes into contact with a flame. The lowest flash temperature is called the flash point value of the oil.
The HTBK-1001 fully automatic open flash point and fire point tester is manufactured in accordance with the standard methods of GB267 “Determination of Flash Point and Fire Point of Petroleum Products (Open Cup Method)” and GB3536 “Determination of Flash Point and Fire Point of Petroleum Products (Cleveland Open Cup Method)”. It is an ideal instrument for open flash point testing in departments such as petroleum, chemical engineering, power, environmental protection, railways, and scientific research.
Main features:
- It adopts a high-speed signal processor for control and an adaptive PID control algorithm to automatically adjust the temperature rise curve according to standard requirements;
- It is equipped with an automatic atmospheric pressure detection chip, which automatically measures, corrects and corrects the flash point value of atmospheric pressure;
- Data storage function, capable of storing 1000 results, allowing for real-time access to historical records;
- Differential detection, automatic correction of system deviation, automatic stop of detection and alarm when temperature exceeds the limit;
- The ignition, detection, alarm, cooling and printing instruments automatically complete the measurement;
- Safe and gas-free, platinum alloy electric heating wire ignition method, safe, stable and reliable, with high testing accuracy;
Main technical indicators:
Display mode | The 10.1-inch touch screen with both Chinese and English displays the operation interface in both languages |
Measurement method | Differential detection formula for thermocouples |
Display range | -59.9 degrees to 400 degrees |
Measurement range | Room temperature -400℃ |
Repeatability | ±4℃ |
Printer | High-speed thermal printer, paper width 56 mm |
Ignition method | Automatic electronic ignition |
Temperature detection | Imported precision platinum resistance |
Self-check function | Printer detection arm lifting and lowering, ignition rod sweeping, etc |
Overall dimensions | 450×280×290mm |
Power supply | AC220V±10% |
Weight | About 15 kg |