Product Purpose:
Equipped with both cooling and heating bidirectional temperature control systems and temperature controllable functions. It is an indispensable laboratory device for research and teaching departments in botany, biology, microbiology, genetics, virology, environmental protection, etc. Widely used for constant temperature experiments, culture testing, environmental testing, etc.
Product Features:
- Acrylic panel, LCD multi-data full-screen operation display, oversized 4.5-inch LCD screen. Uses temperature adaptive control, automatically determines compressor on/off temperature points. High precision, no overshoot at high temperatures. Features power failure memory, power failure compensation, fixed value operation, timed operation, and automatic stop functions.
- Double-layer door structure. After opening the outer door, the working chamber can be observed through the inner glass door without affecting the internal temperature. Built-in LED light tube facilitates night-time sampling.
- Outer shell made of high-quality cold-rolled steel plate with electrostatic spraying. Inner chamber made of stainless steel plate with arc design for easy cleaning.
- Shelves inside the working chamber can be adjusted in height freely.
Main Technical Parameters:
Model | SHX70IV | SHX150IV | SHX250IV | SHX350IV | |
Air Circulation | Forced convection | ||||
Performance | Operating Temp. Range | 5℃-60℃ | |||
Temp. Resolution | 0.1℃ | ||||
Temp. Fluctuation | ±1℃ | ||||
Temp. Uniformity | ±1℃ | ||||
Construction | Inner Chamber | Stainless steel plate | |||
Outer Shell | Cold-rolled steel plate, surface electrostatic spraying | ||||
Insulation | Polyurethane | ||||
Heater | Stainless steel heating tube | ||||
Rated Power (W) | 640 | 740 | 860 | 960 | |
Compressor | Air-cooled hermetic compressor | ||||
Refrigerant | R134A | ||||
Light Tube | Illumination tube for Constant Temp/Humidity Incubator / Germicidal tube for Mold Incubator | ||||
Controller | Defrost System | Automatic | |||
Temp. Control Mode | Single-stage LCD, proportional band control | ||||
Temp. Setting Mode | Touch key setting | ||||
Temp. Display Mode | Measured temp (LCD upper display); Set temp (LCD lower display) | ||||
Timer | 0-9999 minutes/hours | ||||
Operation Functions | Fixed value operation, Timed operation, Auto stop, Program operation. | ||||
Program Mode | Standard | ||||
Sensor | Pt100 | ||||
Additional Functions | Deviation correction, Menu key lock, Power failure compensation, Power failure memory. | ||||
Specifications | Internal Dim. (DWH) mm | 340*400*500 | 400*470*780 | 500*570*850 | 560*615*1020 |
External Dim. (DWH) mm | 640*590*1110 | 700*670*1390 | 800*760*1460 | 860*800*1630 | |
Chamber Volume (L) | 70 | 150 | 250 | 350 | |
Shelf Load Capacity | 15Kg per layer | ||||
Number of Shelves | 2 | 3 | 3 | 3 | |
Power Supply | 220V |